AII GPR-1600 Oxygen Analyzer


Features

  • Measurement range:
    • 0-10, 0-100, 0-1000 ppm, 0-1%
    • 0-25% calibration only
  • High accuracy (±0.02ppm in lowest range)
  • 24 month sensor life with normal usage
  • Sample in CO2 backgrounds with XLT sensor
  • Bench, rack or wall mounting options
  • Two Alarms
  • Integrated bypass valve (for GPR-1600)
  • Easy access and sensor replacement
  • Easy to use HMI

$4,896.00 $5,000.00

  • The AII GPR-1600 Oxygen Analyzer has an integral bypass sample system. The analyzer can be temporarily exposed to ambient air to perform a span calibration (20.9% O2).
  • The AII GPR-1600 is designed for monitoring gas in industrial processes where trace oxygen from low parts per million to pure O2 has to be precisely measured. Utilizing AII’s high performance galvanic oxygen sensors these advanced instruments are simple to use with a common, across-the-range chassis, HMI and menu structure, so an operator only needs to learn operation of just one instrument for multiple gas analysis applications at various oxygen levels.Available as bench, 19in rack, panel and wall mounted for flexible installation.
Category:
Measurement range 0-10, 0-100, 0-1000 ppm, 0-1%
(0-25% calibration only)
Accuracy < ±2% of selected range
Response time T90 < 10 seconds
Recovery time 60 sec in air to < 10 ppm in < 1 hour on N2  purge
Sensitivity (LDL) 50 ppb
Linearity < 0.5% of scale
Sensor model GPR-12-333
XLT-12-333 for gas mixture with > 0.5% CO2
Sensor life at 25°C and 1 atm 24 months in < 1000 ppm O2
Calibration interval Typically:  1-3 months
Inlet pressure 0.34 – 2 barg (5-30 psig) with atmospheric vent
Flow rate (constant) 0.5 – 1.0 Nl/min (1-2 SCFH)
Gas connections 1/4″ compression tube fittings
Display Graphical LCD 12.7 x 7cm (5 x 2.75”); resolution 0.01
Enclosure Painted aluminum
See individual mounting options for dimensions
Compensation Barometric pressure and temperature;
Analog output 4-20 mA isolated, 0-1V, and 0-5V
Range ID 1-5 V or 4-20mA, optional relay contacts
Communications Choose from RS485, RS232 or USB
Alarms Two user-adjustable alarm relays
Operating temperature GPR sensor: 5°C to 45ºC (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)
Power Universal 100-240 V AC