Michell Instruments XZR200 Oxygen Analyzer


Features

  • Measurement ranges:
    • 0-25% (standard)
    • 0-100%
  • Zirconium dioxide measurement technology
  • Accuracy:
    • 0-25%; 1% of reading <0.5% O2
    • 0-100%; <1% of reading
  • Configurable outputs:
    • 4-20 mA
    • 0 to 10 VDC
  • Output resolution: 0.01V, 0.01mA or 0.01% O2
  • Automatic or manual calibration
  • Diecast aluminium case IP65
  • Probe lengths: 8.27″ or 15.75″
  • Supplied with mating connector
  • Sample temperature 752°F (400°C)

$1,320.00$1,515.00

Thumbnail Probe Cable PriceTotalQuantity
Michell Instruments XZR200 Oxygen Analyzer - 8.27 8.27" (210mm) Not included $1,320.00 $1,320.00
spinner
Michell Instruments XZR200 Oxygen Analyzer - 8.27 8.27" (210mm) 5m cable $1,515.00 $1,515.00
spinner
Michell Instruments XZR200 Oxygen Analyzer - 15.75 15.75" (400mm)Not included $1,320.00 $1,320.00
spinner
Michell Instruments XZR200 Oxygen Analyzer - 15.75 15.75" (400mm) 5m cable $1,515.00 $1,515.00
spinner
  • The Michell Instruments XZR200 Oxygen Analyzer is a cost effective zirconium-dioxide analyzer to measure percentage level oxygen in combustion processes, ambient air monitoring and many more applications. The unit is configurable to measure either 0-25% or 0-100% oxygen and offers manual or automatic calibration to suit the customer’s needs. Two probe lengths are available (210mm & 400mm) for greater flexibility.
SKU: N/A Category:
Measurement technology

Zirconium Dioxide

Gas

Oxygen

Measurement range

0-25% or 0-100%

Output resolution

0.01 V, 0.01 mA or 0.01% O2

Accuracy (0-25%)

1% of reading < 0.5% O2

Accuracy (0-100%)

< 1% of reading

Response time (T90)

< 5 seconds

Repeatability

< 0.5%

Sample flow rate

0 to 1.7 m/sec

Sample flow effect

±0.5% of full scale

Sample pressure

Atmospheric

Sample temperature

+250°C or +400°C

Sample cell temperature

+700°C

Background gas

Air, N2, CO2, Ar or He

Power supply

24 V DC, ±10%

Power consumption

500 mA maximum @ 24 V DC

Analog outputs

4-20 mA and 0 to 10 V DC

Output ranges

0-25% or 0-100%

Digital communications

RS232 (not available if 4-20 mA output selected)

Ambient temperature

-10 to +85°C

Warm Up time

Approx. 10 minutes

Stabilization time

Included in the above

Dimensions

52 x 75 x 80mm (h x w x d) excluding probe

Probe dimensions

210 or 400mm (length) ø12mm

Weight

< 0.5kg

Wetted materials

Stainless steel

Process connection

12mm Swagelok® connector

Ingress protection

IP65

Housing material

Waterproof die-cast aluminium housing

Warning

Probe tip gets hot, do not touch without PPE

Probe

8.27" (210mm), 15.75" (400mm)

Cable

Not included, 5m cable